Comparative Performance Analysis of Imported vs. Local NIBP Simulators against Blood Pressure Monitor Using Uncertainty Method

Comparative Performance Analysis of Imported vs. Local NIBP Simulators against Blood Pressure Monitor Using Uncertainty Method

Authors

  • Safariati Isabela Universitas Islam Sultan Agung
  • Eka Nuryanto Budisusila Universitas Islam Sultan Agung

DOI:

https://doi.org/10.59485/jtemp.v6i2.154

Keywords:

Calibration, Non-Invasive Blood Pressure, Measurement Uncertainty

Abstract

Calibration of non-invasive blood pressure (NIBP) devices is essential to ensure measurement accuracy in healthcare services. This study aims to compare the performance of two calibrators, XB1000 and Multisim, in testing two NIBP units using the uncertainty analysis method. Measurements were conducted on systolic, diastolic, and heart rate parameters with a total of 16 data points per parameter. The results show that XB1000 achieved a lower total average expanded uncertainty (U95) of 0.421 mmHg compared to Multisim at 0.529 mmHg. Specifically, XB1000 demonstrated the lowest uncertainty across all parameters, with an overall 20.3% improvement over Multisim. Error analysis indicated that systolic measurements had an average error of 2.063 mmHg, diastolic -1.167 mmHg, and heart rate -0.062 bpm. Based on these findings, XB1000 is recommended as the preferred calibrator for NIBP devices, as it provides lower measurement uncertainty and improved accuracy across all tested parameters.

Downloads

Download data is not yet available.

References

Singh A, Rehman SU, Yongchareon S, Chong PHJ. Multi-Resident Non-Contact Vital Sign Monitoring Using Radar: A Review. IEEE Sens J. 2021 Feb 15;21(4):4061–84.

Eder Y, Eldar YC. Sparsity Based Non-Contact Vital Signs Monitoring of Multiple People Via FMCW Radar. 2022 May 10; Available from: http://arxiv.org/abs/2205.05152

Posthuma LM, Downey C, Visscher MJ, Ghazali DA, Joshi M, Ashrafian H, et al. Remote wireless vital signs monitoring on the ward for early detection of deteriorating patients: A case series. Int J Nurs Stud. 2020 Apr;104:103515.

Hadziqoh N, Pertiwi Y, Mulyadi R. ANALISIS HASIL PENGUJIAN DAN KALIBRASI PARAMETER BLOOD PRESSURE PADA ALAT KESEHATAN BLOOD PRESSURE MONITOR. Journal of Public Health Sciences [Internet]. 2022;(1):45–50. Available from: https://jurnal.stikes-alinsyirah.ac.id/index.php/kesmas

Rahmawati E, Widyowati A, Elina AD. IMPLEMENTATION OF CERDIK BEHAVIOR FOR BLOOD PRESSURE CONTROL AT PETERONGAN HEALTH CENTER. Journal of Indonesian Public Health Service. 2025 Apr 27;2(1):9–19.

Purwandari H. ANALYSIS OF FACTORS INFLUENCING BLOOD PRESSURE IN ELDERLY POSYANDU PARTICIPANTS IN NGANJUK DISTRICT [Internet]. Vol. 1, Journal of Health Sport and Nursing. 2025. Available from: http://jurnal.nersunesa.ac.id

Ulina S, Nazara DP. Analisis Sistem Kerja Dan Pengaruh Non-Invasive Blood Pressure Pada Patient Monitor. Jurnal Mutiara Elektromedik. 2023;7(1):7–12.

Riyan Hapipi L, Arief Hidayat M, Dwi Anggraynie R. RANCANG BANGUN ALAT KALIBRASI DIGITAL PRESSURE METER MENGGUNAKAN ARDUINO UNO. MEDIKA TRADA. 2022 Apr 24;2(1).

Utomo S. Valuasi Teknologi Terhadap Produk Alat Kesehatan Industri Manufaktur Sebagai Upaya Peningkatan Inovasi dan Daya Saing Menggunakan Metode TMCRL (Technology Manufacturing Commercialization Readiness Level). Maret [Internet]. 2022 Mar;7(1):89–99. Available from: http://openjournal.unpam.ac.id/index.php/informatika89

Malau MS, Yulizham Y. ANALISA KALIBRASI ALAT INFANT WARMER MENGGUNAKAN ALAT INCUBATOR ANALYZER. JURNAL MUTIARA ELEKTROMEDIK. 2021 Dec 12;5(2):58–65.

Indonesia R. Instruksi Presiden Republik Indonesia Nomor 2 Tahun 2022 tentang Percepatan Peningkatan Penggunaan Produk Dalam Negeri dan Produk Usaha Mikro, Usaha Kecil, dan Koperasi Dalam Rangka Menyukseskan Gerakan Nasional Bangga Buatan Indonesia pada Pelaksanaan Pe [Internet]. 2022. Available from: https://jdih.setkab.go.id/PUUdoc/176683/Inpres_Nomor_2_Tahun_2022.pdf

Published

2025-12-04

How to Cite

Isabela, S., & Nuryanto Budisusila, E. (2025). Comparative Performance Analysis of Imported vs. Local NIBP Simulators against Blood Pressure Monitor Using Uncertainty Method. MEDIKA TRADA, 6(2), 114–123. https://doi.org/10.59485/jtemp.v6i2.154
Loading...