PRECISION AND ACCURACY ANALYSIS OF THE BIO SYSTEMS MODEL A15 CLINICAL CHEMISTRY ANALYZER IN GLUCOSE MEASUREMENT

PRECISION AND ACCURACY ANALYSIS OF THE BIO SYSTEMS MODEL A15 CLINICAL CHEMISTRY ANALYZER IN GLUCOSE MEASUREMENT

Authors

  • Al Vitto Marcelino Marcelino Institut Kesehatan Dan Teknologi Al Insyirah
  • Nur Teknologi Rekayasa Elektro-Medis Institut Kesehatan Dan Teknologi Al Insyrah

DOI:

https://doi.org/10.59485/jtemp.v6i1.68

Keywords:

Precision, Accuracy, glucose, coefficient of variation

Abstract

Blood glucose examination is one of the methods for diagnosing diabetes. Many people experience delays in glucose testing due to concerns about the accuracy of the testing equipment. To produce high-quality results, clinical laboratories require testing equipment with high accuracy and precision. This study aimed to evaluate the precision and accuracy of the A15 clinical chemistry analyzer in measuring glucose parameters. This was a quantitative descriptive study, involving 30 samples and calculating the precision using the coefficient of variation (CV) formula and accuracy using the d% formula. The results showed that the obtained CV value was 4.93%, below the maximum tolerance limit set by the Indonesian Minister of Health Regulation No. 43 of 2013, which is 5%. The obtained d% value was -0.036, indicating a high level of accuracy. It can be concluded that the A15 clinical chemistry analyzer has good precision and accuracy in measuring blood glucose.

 

Downloads

Download data is not yet available.

References

Ditpui. (2023). Diabetes Penyebab

Kematian Tertinggi di Indonesia: Batasi

dengan Snack Sehat Rendah Gula. Direktorat Pengembangan Usaha

Universitas Gajah Mada. https://ditpui.ugm.ac.id/diabetes-penyebab- kematian-tertinggi-di-indonesia-batasi- dengan-snack-sehat-rendah- gula/#:~:text=Data International Diabetes

Federation

Endrianti, R. (2023). SYSMEXXN-330DILABORATORIUMKLINIKLABORAVerification Method Of The SysmexXN-330 Hematology Analyzer In The. Jurnal

Kesehatan Siliwangi, 04(1), 61-69.

Geto, Z., Getahun, T., Lejisa, T., Tolcha, Y., Bikila, D., Bashea, C., Meles, M., Habtu, W., Ashebir, G., Negasa, B., Sileshi, M., Daniel, Y., Gashu, A., & Challa, F. (2022). Evaluation of Sigma Metrics and WestgardRule Selection and Implementation of Internal

Quality Control in Clinical ChemistryReference Laboratory, Ethiopian PublicHealth Institute. Indian Journal of Clinical

Biochemistry, 37(3), 285-293. https://doi.org/10.1007/s12291-021- 00994-x

Hindayani, A., & Hamim, N. (2022). Akurasi

dan Presisi Metode Sekunder PengukuranKonduktivitas Menggunakan Sel Jones TipeE untuk Pemantauan Kualitas Air Minum. IJCA (Indonesian Journal of Chemical

Analysis), 5(1), 41- 51. https://doi.org/10.20885/ijca.vol5.iss1.art5

Labpedia.net. (2020). Quality Control (QC)

of the Clinical laboratory. Labpedia.Net. https://labpedia.net/quality- control-of-the- clinical-laboratory/

Nabity, M. B., Harr, K. E., Camus, M. S., Flatland, B., & Vap, L. M. (2018). ASVCP guidelines:

Allowable total error hematology. Veterinary Clinical Pathology, 47(1), 9-21. https://doi.org/10.1111/vcp.12583

Permenkes. (2013). permenkes no 43 TH2013. Permennkes No 43 Thn 2013, 1-4.

Praja, R. W. (2022). Analisis PengendalianMutuInternal Pemeriksaan Kimia Klinik denganMetode Six Sigma di LaboratoriumPatologi Klinik RSUD Dr. H. Abdul MoeloekProvinsi Lampung. Analisis PengendalianMutu Internal Pemeriksaan KimiaKlinik DenganMetode Six Sigma Di LaboratoriumPatologi Klinik RSUD Dr. H. Abdul Moeloek Provinsi Lampung, 27(8.5.2017), 2003-2005.

Prasetya, H. R., Muhajir, N. F., &Dumatubun, M. P. I. (2021). PenggunaanSixSigma Pada Pemeriksaan Jumlah Leukosit Di Rsud Panembahan Senopati Bantul. Journal of Indonesian Medical Laboratory and Science(JoIMedLabS), 2(2), 165-174. https://doi.org/10.53699/joimedlabs.v2i2. 72

Putra, M. E., Gizi, J., Masyarakat, F. K., Andalas, U., Studi, P., Mesin, T., & Andalas, U. D. (2022). Akurasi Dan Presisi Alat Ukur Tinggi Badan Digital Untuk Penilaian Status Gizi. Jurnal Endurance, 6(3), 616-621. https://doi.org/10.22216/jen.v6i3.580

Shelbi Asrianti, R. D. (2023). Salah Diagnosis Penyakit Berakibat Fatal, Sebabkan Kecacatan hingga Kematian. Republika.https://ameera.republika.co.id/berita/ryo77 o414/salah-diagnosis-penyakit-berakibatfatal-sebabkan-kecacatan-hingga-kematian

Teshome, M., Worede, A., & Asmelash, D.(2021). Total clinical chemistry laboratory errors and evaluation of the analytical quality control using sigma metric for routine clinical

chemistry tests. Journal of Multidisciplinary Healthcare, 14, 125- 136. https://doi.org/10.2147/JMDH.S286679

Tyas, L. C. (2015). Pemeriksaan Kadar Glukosa Darah Yang Diperiksa Secara

Langsung Dan Ditunda 24 Jam. Sekolah Tinggi Ilmu Kesehatan, 37. http://repo.stikesicmejbg.ac.id/4686/1/PEMERIKSAAN KADAR GLUKOSA DARAH.pdf

Kuspranoto, A Haris.(2024). Rancang Bangun Elektrostimulator dengan Output Tiga Gelombang Berbasis Arduino Mega Pro Mini 2560. DOI: https://doi.org/10.18196/mt.v5i2.15867

Wahyudi, W. (2022). Analisis Motivasi Belajar Siswa Dengan Menggunakan Model Pembelajaran Blended Learning Saat Pandemi Covid-19 (Deskriptif Kuantitatif Di Sman 1 Babadan Ponorogo). KadikmA, 13(1), 68.https://doi.org/10.19184/kdma.v13i1.3132 7

Yudita, F., Purbayanti, D., Ramdhani, F. H., & Jaya, E. (2023). Evaluasi Kontrol Kualitas Pemeriksaan Glukosa Darah di Laboratorium X Palangka Raya. Borneo Journal of

Medical Laboratory Technology, 5(2), 358-365. https://doi.org/10.33084/bjmlt.v5i2.5184

Published

2025-06-04

How to Cite

Marcelino, A. V. M., & Nur. (2025). PRECISION AND ACCURACY ANALYSIS OF THE BIO SYSTEMS MODEL A15 CLINICAL CHEMISTRY ANALYZER IN GLUCOSE MEASUREMENT. MEDIKA TRADA, 6(1), 1–8. https://doi.org/10.59485/jtemp.v6i1.68