PRECISION AND ACCURACY ANALYSIS OF THE BIO SYSTEMS MODEL A15 CLINICAL CHEMISTRY ANALYZER IN GLUCOSE MEASUREMENT
DOI:
https://doi.org/10.59485/jtemp.v6i1.68Keywords:
Precision, Accuracy, glucose, coefficient of variationAbstract
Blood glucose examination is one of the methods for diagnosing diabetes. Many people experience delays in glucose testing due to concerns about the accuracy of the testing equipment. To produce high-quality results, clinical laboratories require testing equipment with high accuracy and precision. This study aimed to evaluate the precision and accuracy of the A15 clinical chemistry analyzer in measuring glucose parameters. This was a quantitative descriptive study, involving 30 samples and calculating the precision using the coefficient of variation (CV) formula and accuracy using the d% formula. The results showed that the obtained CV value was 4.93%, below the maximum tolerance limit set by the Indonesian Minister of Health Regulation No. 43 of 2013, which is 5%. The obtained d% value was -0.036, indicating a high level of accuracy. It can be concluded that the A15 clinical chemistry analyzer has good precision and accuracy in measuring blood glucose.
Downloads
References
Ditpui. (2023). Diabetes Penyebab
Kematian Tertinggi di Indonesia: Batasi
dengan Snack Sehat Rendah Gula. Direktorat Pengembangan Usaha
Universitas Gajah Mada. https://ditpui.ugm.ac.id/diabetes-penyebab- kematian-tertinggi-di-indonesia-batasi- dengan-snack-sehat-rendah- gula/#:~:text=Data International Diabetes
Federation
Endrianti, R. (2023). SYSMEXXN-330DILABORATORIUMKLINIKLABORAVerification Method Of The SysmexXN-330 Hematology Analyzer In The. Jurnal
Kesehatan Siliwangi, 04(1), 61-69.
Geto, Z., Getahun, T., Lejisa, T., Tolcha, Y., Bikila, D., Bashea, C., Meles, M., Habtu, W., Ashebir, G., Negasa, B., Sileshi, M., Daniel, Y., Gashu, A., & Challa, F. (2022). Evaluation of Sigma Metrics and WestgardRule Selection and Implementation of Internal
Quality Control in Clinical ChemistryReference Laboratory, Ethiopian PublicHealth Institute. Indian Journal of Clinical
Biochemistry, 37(3), 285-293. https://doi.org/10.1007/s12291-021- 00994-x
Hindayani, A., & Hamim, N. (2022). Akurasi
dan Presisi Metode Sekunder PengukuranKonduktivitas Menggunakan Sel Jones TipeE untuk Pemantauan Kualitas Air Minum. IJCA (Indonesian Journal of Chemical
Analysis), 5(1), 41- 51. https://doi.org/10.20885/ijca.vol5.iss1.art5
Labpedia.net. (2020). Quality Control (QC)
of the Clinical laboratory. Labpedia.Net. https://labpedia.net/quality- control-of-the- clinical-laboratory/
Nabity, M. B., Harr, K. E., Camus, M. S., Flatland, B., & Vap, L. M. (2018). ASVCP guidelines:
Allowable total error hematology. Veterinary Clinical Pathology, 47(1), 9-21. https://doi.org/10.1111/vcp.12583
Permenkes. (2013). permenkes no 43 TH2013. Permennkes No 43 Thn 2013, 1-4.
Praja, R. W. (2022). Analisis PengendalianMutuInternal Pemeriksaan Kimia Klinik denganMetode Six Sigma di LaboratoriumPatologi Klinik RSUD Dr. H. Abdul MoeloekProvinsi Lampung. Analisis PengendalianMutu Internal Pemeriksaan KimiaKlinik DenganMetode Six Sigma Di LaboratoriumPatologi Klinik RSUD Dr. H. Abdul Moeloek Provinsi Lampung, 27(8.5.2017), 2003-2005.
Prasetya, H. R., Muhajir, N. F., &Dumatubun, M. P. I. (2021). PenggunaanSixSigma Pada Pemeriksaan Jumlah Leukosit Di Rsud Panembahan Senopati Bantul. Journal of Indonesian Medical Laboratory and Science(JoIMedLabS), 2(2), 165-174. https://doi.org/10.53699/joimedlabs.v2i2. 72
Putra, M. E., Gizi, J., Masyarakat, F. K., Andalas, U., Studi, P., Mesin, T., & Andalas, U. D. (2022). Akurasi Dan Presisi Alat Ukur Tinggi Badan Digital Untuk Penilaian Status Gizi. Jurnal Endurance, 6(3), 616-621. https://doi.org/10.22216/jen.v6i3.580
Shelbi Asrianti, R. D. (2023). Salah Diagnosis Penyakit Berakibat Fatal, Sebabkan Kecacatan hingga Kematian. Republika.https://ameera.republika.co.id/berita/ryo77 o414/salah-diagnosis-penyakit-berakibatfatal-sebabkan-kecacatan-hingga-kematian
Teshome, M., Worede, A., & Asmelash, D.(2021). Total clinical chemistry laboratory errors and evaluation of the analytical quality control using sigma metric for routine clinical
chemistry tests. Journal of Multidisciplinary Healthcare, 14, 125- 136. https://doi.org/10.2147/JMDH.S286679
Tyas, L. C. (2015). Pemeriksaan Kadar Glukosa Darah Yang Diperiksa Secara
Langsung Dan Ditunda 24 Jam. Sekolah Tinggi Ilmu Kesehatan, 37. http://repo.stikesicmejbg.ac.id/4686/1/PEMERIKSAAN KADAR GLUKOSA DARAH.pdf
Kuspranoto, A Haris.(2024). Rancang Bangun Elektrostimulator dengan Output Tiga Gelombang Berbasis Arduino Mega Pro Mini 2560. DOI: https://doi.org/10.18196/mt.v5i2.15867
Wahyudi, W. (2022). Analisis Motivasi Belajar Siswa Dengan Menggunakan Model Pembelajaran Blended Learning Saat Pandemi Covid-19 (Deskriptif Kuantitatif Di Sman 1 Babadan Ponorogo). KadikmA, 13(1), 68.https://doi.org/10.19184/kdma.v13i1.3132 7
Yudita, F., Purbayanti, D., Ramdhani, F. H., & Jaya, E. (2023). Evaluasi Kontrol Kualitas Pemeriksaan Glukosa Darah di Laboratorium X Palangka Raya. Borneo Journal of
Medical Laboratory Technology, 5(2), 358-365. https://doi.org/10.33084/bjmlt.v5i2.5184
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2025 MEDIKA TRADA : Journal of Electromedical Engineering Polbitrada

This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.