Analysis of Accuracy flowrate and Uncertainty of Domestic Syringe Pump Products Using the ISO GUM Method
DOI:
https://doi.org/10.59485/jtemp.v6i2.160Keywords:
Syringe Pump, Accuracy, Measurement Uncertainty, ISO GUM, TKDNAbstract
This study evaluates the accuracy of locally manufactured syringe pumps used in Karanganyar District Hospital and analyzes their measurement uncertainty based on the ISO GUM method. Five Indonesian brands (Endo, Esa Tech, Inmedik, KalTech, and Krisma) were tested using an Infusion Device Analyzer at three flow-rate set-points (10, 50, and 100 ml/h) and an occlusion test at 100 ml/h. All devices met the standard tolerance limits (±10% and <20 psi). The expanded uncertainty ranged from ±0.09 to ±0.70 ml/h, indicating reliable measurement performance. KalTech and Inmedik demonstrated the most stable accuracy, while Esa Tech tended to over-read at low flow rates and Krisma showed significant under-reading at high flow rates, though still within acceptable limits. Statistical tests confirmed no significant differences among brands. Overall, the findings support that Indonesia-made syringe pumps meet technical requirements and strengthen the implementation of the national TKDN policy for reducing dependence on imported medical devices.
Downloads
References
Heryani N, Purwaningsih E, Rumah Sakit A. Analisis Manajemen Logistik Alat Medis di RSUD Inche Abdoel Moeis Samarinda. 2024.
Dian Irmawati, Nani Yuniar, Rahman. Implementasi pengelolaan alat medis di puskesmas wua-wua kota kendari tahun 2022. Jurnal administrasi dan kebijakan kesehatan. 2023 Apr 1;4.
Medical WHO. HUMAN RESOURCES The role of biomedical engineers WHO Medical device technical series.
Vieira NM, Pires MP de O, Crespo GB, Nascimento LPP, Peterlini MAS, Pedreira MLG. Start-up delay in syringe infusion pumps with different rates and priming techniques of intravenoust sets. Rev Gaucha Enferm. 2022;43.
International C, Commission E, Office IECC, Box PO. International Electrotechnical Commission [IEC], 2012. 2012;
Lasiyah N, Romi Mulyadi, Okta Nola Yolanda iriani. Analisis Manajemen Risiko Keamanan Dan Kualitas Penggunaan Peralatan Medis Di Rumah Sakit Umum Daerah Arifin Achmad Provinsi Riau. Al-Tamimi Kesmas: Jurnal Ilmu Kesehatan Masyarakat (Journal of Public Health Sciences). 2024 Jul 15;13(1):159–67.
Huda Choirul. Kalibrasi Infusion Pump & Syringe Pump. Kementrian Kesehatan Republik Indonesia. 2024;
Laia A, Yulizham Y. Analisis Kalibrasi Syringe Pump. Jurnal Mutiara Elektromedik. 2022 Jun 26;2022(1):25–32.
Ika Santi Widyasari, Tito Yustiawan. Manajemen Peralatan Kesehatan Klinik Medical Center Ptn Di Jawa Timur [Internet]. 2020 Mar. Available from: http://e-journal.unair.ac.id/JPHRECODE
Permenperin No 31 2022. Permenperin No 31 tahun 2022. Peraturan Menteri Perindustrian Republik Indonesia Nomor 31 Tahun 2022. 2022;
Nova LS, Puspitasari DA, Febriani F. Penerapan Kebijakan Tingkat Komponen Dalam Negeri (TKDN) Terhadap Pengembangan Industri Alat Kesehatan. Jurnal Bidang Ilmu Kesehatan. 2025 Sep 30;15(3):231–9.
Hariyono MA, Erlita UA, Wibowo BS, Persadha G, Japeri J, Yakub S, et al. Pelayanan Kesehatan Melalui Standarisasi Peralatan Kesehatan Di Puskesmas Barabai Kabupaten Hulu Sungai Tengah. Lumbung Inovasi: Jurnal Pengabdian kepada Masyarakat. 2023 Jun 11;8(2):143–53.
Dia Sari N, Hubban Fathia M, Nur Syahidah H, Rahmadani Sagala, Zanzabila Aulya. ANALISIS PEMELIHARAAN ALAT KESEHATAN DAN PELAYANAN DI DINAS KESEHATAN KABUPATEN BATUBARA. 2025 Jun;6.
Dabukke Hotromasar, Adiansyah, Sijabar Salomo, Panjaitan Beerkat. Sosialisasi Kalibrasi dan Pemeliharaan Alat Patien Monitor. Journal Abdimas Mutiara. 2025 Feb 7;6:302–6.
Malau MS, Yulizham Y. ANALISA KALIBRASI ALAT INFANT WARMER MENGGUNAKAN ALAT INCUBATOR ANALYZER. JURNAL MUTIARA ELEKTROMEDIK. 2021 Dec 12;5(2):58–65.
Ungusari E. Permenkes no 54 tahun 2015 tentang pengujian dan kalibrasi. 2015;151:10–7.
Kemenkes. Permenkes no 54 tahun 2015 tentang pengujian dan kalibrasi. 2008;1969(1):1–24.
Saguni A. Metode Kerja dan Kalibrasi Alat Kesehatan. Vol. 70. 2018.
Williams JH. Guide to the Expression of Uncertainty in Measurement (the GUM). In: Quantifying Measurement. 2016. p. 6-1-6–9.
JCGM_100_2008. Evaluation of measurement data — Guide to the expression of uncertainty in measurement. 2008;(September).
Hadziqoh N, Pertiwi Y, Mulyadi R. Analisis Hasil Pengujian Dan Kalibrasi Parameter Blood Pressure Pada Alat Kesehatan Blood Pressure Monitor. Journal of Public Health Sciences [Internet]. 2022;(1):45–50. Available from: https://jurnal.stikes-alinsyirah.ac.id/index.php/kesmas
Badan Standarisasi Nasional. SNI IEC 60601-2-24-2016 pompa infus dan pengontrolnya_09869BAR.pdf. 2016.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2025 MEDIKA TRADA

This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.